64 . Lambda 365 Users Guide
D2 Lamp Replacement
1. Make sure that the main power is turned off. If the system and lamp(s) were turned
on at least 20 minutes should pass before the lamp changing process is started. This
time allows the lamps to cool and return to room temperature.
CAUTION
ATTENTION
Use appropriate disposable gloves during the lamp replacement
process. Also avoid touching the glass bulb
section of the lamp this can
leave oil deposits on the bulb which can drastically reduce lamp life.
Utiliser des gants jetables appropriés pendant le processus de
remplacement de la lampe. Evitez également de toucher la section de
l'ampoule de verre de la lampe, cela peut laisser des dépôts d'huile sur
le bulbe qui peut réduire considérablement la vie de la lampe.
2. Remove the white bolt covers (four each) using screwdriver and remove the M3*12
screws (four each) using a Phillips screwdriver to open the vent cover.
3. Remove the two M3*6 screws on the sub cover using a Phillips screwdriver.

Other manuals for PerkinElmer Lambda 365

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PerkinElmer Lambda 365 and is the answer not in the manual?

PerkinElmer Lambda 365 Specifications

General IconGeneral
BrandPerkinElmer
ModelLambda 365
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals