Safety Practices . 29
Fuses
WARNING
There is risk of receiving a fatal electric shock if the fuses are replaced
with the power cord connected.
Il y a un risque d’électrocution si les fusibles sont remplacés tandis que
le cordon d'alimentation est encore branché.
The fuse is located at the rear of the instrument.
1. Turn off and unplug the instrument.
2. Carefully open the compartment latch where the fuse is located.
3. Disconnect the fuse.
4. Replace with two new T3.15L fuses, one for each phase.
5. Close the compartment door.
6. Plug in the instrument and turn on.
Location of latch
on the fuse
compartment
door