Page #45 background image
- 41 -
Troubleshooting
Problem Why it happened What to do
Nothing
happens
when you
apply power
to the device.
There’s no power
at the outlet or
the device is
unplugged.
If you are using AC power, check the outlet and verify that
the device is properly plugged in. Make sure there is power
available at the outlet. Make sure the AC power cord is
connected correctly to the power supply and the power supply
cord is securely connected to the device’s power inlet. If the
problem continues to occur, contact your home care provider.
Return both the device and power supply to your provider, so
they can determine if the problem is with the device or power
supply.
If you are using DC power, make sure your DC power cord and
battery adapter cable connections are secure. Check your
battery. It may need recharged or replaced. If the problem
persists, check the DC cord’s fuse following the instructions
supplied with your DC cord. The fuse may need to be replaced.
If the problem still occurs, contact your home care provider.
The airow
does not turn
on.
There may be a
problem with the
blower.
Make sure the device is powered correctly. Press the Therapy
button on top of the device to start airow.
If the airow does not turn on, there may be a problem with your
device. Contact your home care provider for assistance.
The devices
display is
erratic.
The device has
been dropped or
mishandled, or
the device is in an
area with high
Electromagnetic
Interference (EMI)
emissions.
Unplug the device. Reapply power to the device. If the problem
continues, relocate the device to an area with lower EMI
emissions (away from electronic equipment such as cellular
phones, cordless phones, computers, TVs, electronic games, hair
dryers, etc.). If the problem still occurs, contact your home care
provider for assistance.

Table of Contents

Other manuals for Philips DreamStation 2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips DreamStation 2 and is the answer not in the manual?

Philips DreamStation 2 Specifications

General IconGeneral
Pressure Range4-20 cm H2O
HumidificationIntegrated heated humidifier
Heated TubeOptional
DisplayColor Touchscreen
Altitude CompensationAutomatic
Device TypeCPAP
Ramp Time0 to 45 min (5-minute increments)
Flex Pressure ReliefA-Flex
Power Supply100-240V, 50/60Hz
ConnectivityBluetooth
Data StorageSD card, Cloud
Weight2.5 lbs (without humidifier)
Sound LevelLess than 27 dBA