- 53 -
Heater plate Max Temperature: 68°C (15F)
Humidity Humidity
min
Output: 12 mg H
2
O/L
Measured at expected leak across therapy pressures , 17 °C– 35°C,
15% RH for all compatible tubes.
Maximum delivered
gas temperature
<43°C (109.4°F)
Declared dual-number noise emissions values (in accordance with ISO 4871)
Tube Size Sound Pressure
Level
Uncertainty Sound Power
Level
Uncertainty
15 and 22 (mm)
tubing type
27 dB(A) 2 dB(A) 35 dB(A) 2 dB(A)
12 (mm) tubing
type
27 dB(A) 2 dB(A) 35 dB(A) 2 dB(A)
Pressure accuracy
Pressure Increments: 4-20 cmH
2
O (in 0.5 cmH
2
O increments)
Maximum static and dynamic pressure accuracy, according to ISO 80601-2-70:2015:
Tube Type Static Dynamic
4 to 20 cmH
2
O
15 and 22 (mm) tubing type ± 0.5 cmH
2
O ± 1.0 cmH
2
O
12 (mm) tubing type ± 1.0 cmH
2
O ± 2.0 cmH
2
O
Static pressure accuracy has a measurement uncertainty of 1.31%.
Dynamic pressure accuracy has a measurement uncertainty of 2.70%.

Table of Contents

Other manuals for Philips DreamStation 2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips DreamStation 2 and is the answer not in the manual?

Philips DreamStation 2 Specifications

General IconGeneral
BrandPhilips
ModelDreamStation 2
CategoryRespiratory Product
LanguageEnglish

Related product manuals