110-240 12
E. Replace the reservoir cap, open the inlet and outlet valves, and restart the Chiller.
F. Observe the liquid level in the external reservoir; adjust the valve on the Chiller outlet as
required to maintain a stable fluid level.
CAUTION: When running an open loop system for extended periods, the fluid level in the
Chiller reservoir should be checked periodically to avoid low fluid conditions.
To check the reservoir fluid level, close the inlet and outlet valves, turn the Chiller ‘off’, and remove
the reservoir cap. Slowly open the inlet and outlet valves and allow fluid to drain from the external
reservoir into the Chiller reservoir. Close the valves when the fluid level within the Chiller reservoir
reaches the top of the filler neck. Add fluid to the external reservoir as required. Replace the
reservoir cap, open the inlet and outlet valves, and turn the Chiller back on.