72
Command – 1 ÷ 3 marks
stability marker – [space] if stable
[?] if unstable
[^] if there is an error of exceeding the range on +
[v] if there is an error of exceeding the range on -
mark – [space] for positive values or [-] for negative values
mass – 9 marks alignment to the right
unit – 3 marks alignment to the left
16. WEIGHING LOADS UNDER BALANCE
In standard analytical and precision balances can weigh load on suspension.
To use this function:
• Remove plastic plug in basic of balance
• There is suspension in the basic of balance. It’s installed for good.
• Install hook to hang load (hook isn’t standard equipment for balance), weigh
load on hook.
Attention:
• Suspension cannot be turned around, move or manipulate. Mechanism of
balance can be damaged this way.
• Mass all additional elements as balance, string should be settled to zero
by pressing key Esc/TARE.
17. LIST OF ORDERS IN RADWAG BALANCES
Function TARA
Format T CR LF (setting to tare)
Function ZERO
Format Z CR LF (setting to zero)
Function PRINT
Format SI CR LF (sending results when stab)