XCell
TM
ATF System with C410:V4B Controller User Guide
108
OPC Inputs
Tagname plc in DB41
PLC Address
Definition
Units
Data
type
Structure
Tagname in structure
"DB41".
ATF_START
%DB41.DBX0.0
Command Word - ATF Start
Bool
"DB41".
ATF_STOP
%DB41.DBX0.1
Command Word - ATF Stop
Bool
"DB41".
AFT_Ack_Alarms
%DB41.DBX0.2
Command Word - Acknowledge Alarms
Bool
"DB41".
Reset_Counters
%DB41.DBX0.3
Command Word - Algorithm Performance Counters
Reset
Bool
"DB41".
ATF_Control_Mode
%DB41.DBX0.4
Command Word - ATF - Control Mode (Flow/Time)
Bool
"DB41".
ATF_Slope_Enabled
%DB41.DBX0.5
Command Word - ATF - Slope Enable
Bool
"DB41".
CommandApply
%DB41.DBX0.6
Command Word - Command Apply
Bool
"DB41".
ConfigSave
%DB41.DBX0.7
Command Word - Configuration Save
Bool
"DB41".
ATF_ConnectedBR
%DB41.DBW2
Connected Bioreactor (INT number)
Int
"DB41".
PFLOWSP
%DB41.DBD4
Pressure Flow Setpoint LPM
LPM
Real
"DB41".
PTIMESP
%DB41.DBD8
Pressure Flow Setpoint SEC
SEC
Real
"DB41".
EFLOWSP
%DB41.DBD12
Exhaust Flow Setpoint LPM
LPM
Real
"DB41".
ETIMESP
%DB41.DBD16
Exhaust Flow Setpoint SEC
SEC
Real
"DB41".
PRV1
%DB41.DBD20
Flow Regulator Setpoint (bar)
bar/PSI
Real
"DB41".
PRV2
%DB41.DBD24
Exhaust Regulator Setpoint (bar)
bar/PSI
Real
"DB41".
PV1
%DB41.DBD28
Flow Prop Valve Setpoint (%)
%
Open
Real
"DB41".
PV2
%DB41.DBD32
Exhaust Prop Valve Setpoint (%)
%
Open
Real
"DB41".
BatchName[0]
%DB41.DBB36
Batch name, 5 Char (bytes)
Char
"DB41".
BatchName[1]
%DB41.DBB37
Batch name, 5 Char (bytes)
Char
"DB41".
BatchName[2]
%DB41.DBB38
Batch name, 5 Char (bytes)
Char
"DB41".
BatchName[3]
%DB41.DBB39
Batch name, 5 Char (bytes)
Char
"DB41".
BatchName[4]
%DB41.DBB40
Batch name, 5 Char (bytes)
Char
"DB41".
Model_Num_HMI
%DB41.DBW42
Model Number SP
Int
"DB41".
WATSP
%DB41.DBW44
Warning/Alarm Acknowledge Delay Time SP
Min
Int
"DB41".
Bioreactor_Backpressure
%DB41.DBD46
Bioreactor Backpressure
bar/PSI
Real
"DB41".
Height_Differential
%DB41.DBW50
Height Differential SP
CM
Int
"DB41".
FDFOFST
%DB41.DBD52
Flow Driving Force Offset for switching
bar/PSI
Real
"DB41".
Press_Cycle_Delay_Time_SP
%DB41.DBW56
Pressure Cycle Delay Time SP
Stroke
%
Int
"DB41".
Press_Cycle_Over_Time_SP
%DB41.DBW58
Pressure Cycle Over Time SP
Stroke
%
Int
"DB41".
Pres_Cycl_PV_Step_Size
%DB41.DBD60
Pressure Cycle - PV Step Size - %
Stroke
%
Real
"DB41".
EDFOFST
%DB41.DBD64
Exhaust Driving Force Offset for switching
bar/PSI
Real
"DB41".
Vac_Cycle_Delay_Time_SP
%DB41.DBW68
Vacuum Cycle Delay Time SP
Stroke
%
Int
"DB41".
Vac_Cycle_Over_Time_SP
%DB41.DBW70
Vacuum Cycle Over Time SP
Stroke
%
Int
"DB41".
Exhs_Cycl_PV_Step_Size
%DB41.DBD72
Exhaust Cycle - PV Step Size - %
Stroke
%
Real
"DB41".
P2_Pressure_Alarm_Hi_Hi_SP
%DB41.DBD76
P2 Alarm Limit Hi Hi Set Point
bar/PSI
Real
"DB41".
P2_Alarm_Hi_Hi_TMR_SP
%DB41.DBW80
P2 Pressure Alarm Hi Hi - Timer SP
SEC
Int
"DB41".
P2_Pressure_Alarm_Lo_Lo_SP
%DB41.DBD82
P2 Alarm Limit Lo Lo Set Point
bar/PSI
Real
"DB41".
P2_Alarm_Lo_Lo_TMR_SP
%DB41.DBW86
P2 Pressure Alarm Lo Lo - Timer SP
SEC
Int
"DB41".
PFLOWMAXSP
%DB41.DBD88
Max Pressure Flow Setpoint LPM
LPM
Real
"DB41".
PFLOWMINSP
%DB41.DBD92
Min Pressure Flow Setpoint LPM
LPM
Real
"DB41".
EFLOWMAXSP
%DB41.DBD96
Max Exhaust Flow Setpoint LPM
LPM
Real
"DB41".
EFLOWMINSP
%DB41.DBD100
Min Exhaust Flow Setpoint LPM
LPM
Real
"DB41".
PTIMEMAXSP
%DB41.DBD104
Max Pressure Flow Setpoint SEC
SEC
Real
"DB41".
PTIMEMINSP
%DB41.DBD108
Min Pressure Flow Setpoint SEC
SEC
Real
"DB41".
ETIMEMAXSP
%DB41.DBD112
Max Exhaust Flow Setpoint SEC
SEC
Real
"DB41".
ETIMEMINSP
%DB41.DBD116
Min Exhaust Flow Setpoint SEC
SEC
Real
"DB41".
PV1MAX
%DB41.DBD120
Flow Prop Valve Max Value (%)
%
Open
Real
"DB41".
PV1MIN
%DB41.DBD124
Flow Prop Valve Min Value (%)
%
Open
Real
"DB41".
PV2MAX
%DB41.DBD128
Exhaust Prop Valve Max Value (%)
%
Open
Real
"DB41".
PV2MIN
%DB41.DBD132
Exhaust Prop Valve Min Value (%)
%
Open
Real
"DB41".
PRV1MAX
%DB41.DBD136
Flow Regulator Max Value (bar)
bar/PSI
Real
"DB41".
PRV1MIN
%DB41.DBD140
Flow Regulator Min Value (bar)
bar/PSI
Real
"DB41".
PRV2MAX
%DB41.DBD144
Exhaust Regulator Max Value (bar)
bar/PSI
Real
"DB41".
PRV2MIN
%DB41.DBD148
Exhaust Regulator Min Value (bar)
bar/PSI
Real
Light Blue highlighted inputs are activated when CommandApply bit is set
Light Red highlighted inputs are activated when ConfigurationSave bit is set

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the REPLIGEN XCell ATF Series and is the answer not in the manual?

REPLIGEN XCell ATF Series Specifications

General IconGeneral
BrandREPLIGEN
ModelXCell ATF Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals