XCell
TM
ATF System with C410:V4B Controller User Guide
71
Appendix 3: Controller Lists: Alarm, Warning, Inputs & Outputs
Alarms/Warning List
Tag Name
Description
Alarm
Enable
Condition
Delay
Action
"DB1".ALARM_REG(0)
XCell™ ATF
Warning condition
has not been reset
within the alarm
delay time, XCell™
ATF function halted.
Pump is
Running and
Alarm Delay
Timer Setpoint
is Greater than
Zero.
*See Note
Number of
minute(s).
Determined by
Alarm Delay
parameter on
Basic Set Up
Screen
Alarm,
System Stop,
Activate System
Fault Warning Horn
"DB1".ALARM_REG(1)
Main Power Loss
(120/220) while
Pump was running
First Scan of
PLC on Power
Up.
On first scan Pump
State is not 0
N/A
Alarm,
System Stop,
Activate System
Fault Warning Horn
"DB1".ALARM_REG(2)
P2 Pressure Above
HI HI Limit
Pump is
Running
P2 pressure is
greater than or
equal to the P2 Hi
HI Pressure alarm
setpoint.
P2 Hi Hi Set
Point Alarm
Delay in 10ms
increments.
Alarm,
System Stop,
Activate System
Fault Warning Horn
"DB1".WARN_REG(0)
Flow Set Point
cannot be reached.
P-Flow Regulator
(PV1) above
maximum operating
setting
Pump is
Running
PV1 >= PV1MAX
N/A
Warning,
Activate System
Fault Warning Horn
"DB1".WARN_REG(1)
Flow Set Point
cannot be reached.
P-Flow Regulator
(PV1) below
minimum operating
setting
Pump is
Running
PV1 <= PV1MIN
N/A
Warning,
Activate System
Fault Warning Horn
"DB1".WARN_REG(2)
Exhaust Set Point
cannot be reached.
E-Flow Regulator
(PV2) above
maximum operating
setting
Pump is
Running
PV2 >= PV2MAX
N/A
Warning,
Activate System
Fault Warning Horn
"DB1".WARN_REG(3)
Exhaust Set Point
cannot be reached.
E-Regulator (PV2)
below minimum
operating setting
Pump is
Running
PV2 <= PV2MIN
N/A
Warning,
Activate System
Fault Warning Horn
"DB1".WARN_REG(4)
P2 Pressure below
Lo Lo Limit
Pump is
Running
P2_PV <=
Pressure Alarm Lo
Lo Setpoint
P2 Lo Lo Set
Point Alarm
Delay in 10ms
increments.
Warning,
Activate System
Fault Warning Horn
"DB1".WARN_REG(5)
System Stop
Pressed
Always
A Blue system stop
buttons are
pressed
N/A
Warning, Activate
System Fault
Warning
*NOTE: "DB1".WARN_REG(0) is true, or "DB1".WARN_REG(1) is true, or "DB1".WARN_REG(2) is true, or
"DB1".WARN_REG(3) is true, or "DB1".WARN_REG(4) is true.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the REPLIGEN XCell ATF Series and is the answer not in the manual?

REPLIGEN XCell ATF Series Specifications

General IconGeneral
BrandREPLIGEN
ModelXCell ATF Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals