53
7. After LOGON, the main screen is displayed. Proceed to the SET UP screen.
8. Enter settings as in section 6.4.
Note XCell™ ATF System selection - Select XCell™ ATF 6 System. Note reactor pressure and ∆
height- enter 0 and measured difference between Diaphragm Pump clamp and vessel liquid
level in cm.
9. Power up compressed Air and Vacuum services, confirm pressure and vacuum
services are on by observing the respective pressure gauges.
10. Connect pneumatic Pump Line to Diaphragm Pump air filter.
11. Activate the P2 trend on the MAIN screen. (Any setting adjustments in the P2 screen
are made from the P2 TREND screen).
12. Press the START button on the MAIN screen.
13. XCell™ ATF System should begin cycling normally.
14. Observe for conditions described at the end of the last section.
Note the air purge from the system after start of XCell™ ATF System. Note also the small up and
down change in vessel liquid level, indicating XCell™ ATF System cyclic flow.
Changing flow rate:
To change flow rates, simply go to the main screen select P-Flow SP. Enter the new flow rate on
the pup-up menu. Press Yes to accept current PV and PRV values or press Default to accept
default PV and PRV values. Default is recommended when entering new flow rates. Follow
similar procedures to change E-Flow SP. It is recommended to STOP the controller prior to
change of flow rate. Following the change, re-START the controller. It will start with the
diaphragm in the fully exhausted position.
Filtrate or Harvest:
Start the Filtrate (or Harvest) pump only after XCell™ ATF System flow has stabilized. Stop the
Filtrate / Harvest pump when stopping the XCell™ ATF System flow; accordingly, the C410v4B
controller provides relay outputs on the E-BOX for activating or deactivating a filtrate / harvest
pump.
Please contact your local Repligen Account Manager to obtain additional technical assistance
with automation/integration related information.