XCell
TM
ATF System with C410:V4B Controller User Guide
2
A typical configuration of the XCell™ ATF System is shown in Figure 1. The Filter Housing accepts
either HFM, with pore sizes from 750kD to ~0.2 micron, or a SM for fractionation of larger
particles, >70 microns. The separating element, the HFM or SM, is positioned between a process
vessel or a bioreactor at one end and the Diaphragm Pump at the other end. The vessel serves as
a storage container for the content to be filtered. The Diaphragm Pump provides the means for
generating alternating tangential flow (ATF), moving the contents of the vessel back and forth,
between the vessel and pump, through the hollow fibers of the HFM or through the SM. The
XCell™ ATF process provides the means for generating rapid, low shear, tangential flow, allowing
for retention of the larger components (i.e.- cells) and filtration of smaller components (i.e.
media components).
A filtrate pump as shown in Figure 1 is used for controlled removal of a filtered stream. The
unfiltered material remains in the system. Only a single connection is required between the
XCell™ ATF System and the vessel. As shown in Figure 1, the connection can be through a side
port (or bottom port) commonly configured on large scale stainless steel or single-use
bioreactors, or through the head plate as typical with smaller bioreactors. When placed next to
the vessel, only a short connection, commonly referred to as the A2B Connection, is required
between the XCell™ ATF system and the vessel. This connection can be hard piped or soft piped
and is made in a sterile manner. The filtration process remains closed and therefore sterility
between the vessel and the XCell™ ATF System is maintained.
Figure 1. Filtration Assembly Connection to C410v4B Controller and a Bioreactor Side Port
Note: The XCell™ ATF C410v4B Controller has one additional enclosure, the power box (P-box)
which houses the high voltage electrical components (not displayed here). Please reference the
XCell™ ATF Single-use Start up Guide for proper single-use connectivity.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the REPLIGEN XCell ATF Series and is the answer not in the manual?

REPLIGEN XCell ATF Series Specifications

General IconGeneral
BrandREPLIGEN
ModelXCell ATF Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals