Chapter 2 Performance Verification Test RIGOL
DSA1000A/DSA1000 Performance Verification Guide
Chapter 2 Performance Verification Test
This chapter introduces the performance verification test method and procedures of DSA1000A and
DSA1000 series spectrum analyzer.
Note:
1) Make sure the spectrum analyzer has passed the self-calibration before executing the
performance verification test.
2) Make sure that the instrument has been warmed up for at least 30 minutes before executing any
of the following tests.
3) Press Preset to reset the instrument to the factory setting before or after executing any of the
following tests.
4) For amplitude-related tests, perform self-calibration on the instrument (press System
Calibrate Cal Now) before the test.
5) Definitions of “Typical Value” and “Nominal Value” for this product.
Typical Value: the performance specification of this product under specified conditions.
Nominal Value: the approximate quantity of the product during application.