Chapter 2 Functions of the Front Panel of GPSA RIGOL
RSA3000 User Guide 2-73
Marker To
Sets the other system parameters (such as center frequency and reference level) by
using the current marker values. Press any key under Marker To to activate a
marker automatically if no marker is currently enabled.
Mkr->CF
Sets the center frequency of the analyzer to the frequency of the current marker.
⚫ If Position marker is selected, the center frequency will be set to the frequency
of the current marker.
⚫ If Delta marker is selected, the center frequency will be set to the frequency of
the Delta marker.
⚫ This function is invalid in zero span.
Mkr->CF Step
Sets the center frequency step of the analyzer to the frequency of the current
marker.
⚫ If Position marker is selected, the center frequency step will be set to the
frequency of the current marker.
⚫ If Delta marker is selected, the center frequency step will be set to the
frequency difference between the Delta marker and the reference marker.
⚫ This function is invalid in zero span.
Mkr->Start
Sets the start frequency of the analyzer to the frequency of the current marker.
⚫ If Position marker is selected, the start frequency will be set to the frequency of
the current marker.
⚫ If Delta marker is selected, the start frequency will be set to the frequency of the
Delta marker.
⚫ This function is invalid in zero span.
Mkr->Stop
Sets the stop frequency of the analyzer to the frequency of the current marker.
⚫ If Position marker is selected, the stop frequency will be set to the frequency of
the current marker.
⚫ If Delta marker is selected, the stop frequency will be set to the frequency of the
Delta marker.
⚫ This function is invalid in zero span.