Abdomen
AbdomenFast/
AbdomenFast06s
Indications:
Fast spiral mode for abdominal studies, e.g. when a
patient has difficulty with breathhold.
For the SOMATOM Emotion 6-slice configuration:
A range of 20 cm will be covered in 5.64 sec. with
AbdomenFast06s.
Emotion 6 AbdFast
kV 130
Effective mAs/
Quality ref. mAs
120
Rotation time 0.8/0.6 sec.
Acquisition 6 x 3.0 mm
Slice collimation 3.0 mm
Slice width 6.0 mm
Feed/Rotation 27.0 mm
Pitch Factor 1.50
Increment 6.0 mm
Kernel B41s
CTDI
Vol
12.24 mGy
Effective dose Male: 5.34 mSv
Female: 6.91 mSv