Respiratory Gating
Resp
Indications:
A spiral mode for patients with a respiration rate per
minute greater than 10 using respiration gating.
For SOMATOM Emotion 16-slice configuration:
A scan range of 30 cm will be covered in 94.95 sec.
For SOMATOM Emotion 6-slice configuration:
A scan range of 30 cm will be covered in 86.04 sec.
Emotion 16 Resp 2
nd
reconstr.
kV 130
Effective mAs/
Quality ref. mAs
320
Rotation time 0.6 sec.
Acquisition 16 x 1.2 mm
Slice collimation 1.2 mm
Slice width 3.0 mm 3.0 mm
Feed/Rotation 1.9
Pitch Factor 0.1
Increment 2.1 mm 2.1 mm
Kernel B31s B31s
CTDI
vol
35.84 mGy