Respiratory Gating
RespLowBreathRate
Indications:
A spiral mode for patients with a respiration rate per
minute greater than 6.5, using respiration gating.
For SOMATOM Emotion 16-slice configuration:
A scan range of 18 cm will be covered in 87.23 sec.
For SOMATOM Emotion 6-slice configuration:
A scan range of 11 cm will be covered in 93.67 sec.
Emotion 16 RespLow 2
nd
reconstr.
kV 130
Effective mAs/
Quality ref. mAs
320
Rotation time 1.0 sec.
Acquisition 16 x 1.2 mm
Slice collimation 1.2 mm
Slice width 3.0 mm 3.0 mm
Feed/Rotation 2.1
Pitch Factor 0.11
Increment 2.1 mm 2.1 mm
Kernel B31s B31s
CTDI
vol
35.84 mGy