Respiratory Gating
Scanning Information
The expected benefits in radiation treatment planning,
e.g. for lung and abdominal tumors are:
• Information about tumor motion in 3D coordinates
and over time
• More accurate tumor shape delineation and there-
fore a more precise RTP (radiation therapy planning)
• Potential for sparing of healthy tissue, minimization
of PTV (planned target volume) and less side effects
• Potential for dose acceleration and higher cure rate
To minimize motion artifacts, two requirements are
mandatory for a CT system:
– Fast gantry rotation to raise the temporal resolution
for artifact free images
– Prospective triggering of image acquisition in a
sequential mode or retrospective gating of image
reconstruction in a spiral mode based on the record-
ing of the respiratory curve in order to obtain images
during inspiration and expiration phases.
Scan Parameters
A respiratory gated lung spiral with cone correction
will be provided. To be able to work also with low
breathing rates, a fixed pitch factor of a min. of 0.1 is
needed. Slice widths and collimation are the same as
for cardio spiral.
In the Specials folder different scan protocols are pre-
defined for different breathing rates.

Table of Contents

Other manuals for Siemens SOMATOM

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens SOMATOM and is the answer not in the manual?

Siemens SOMATOM Specifications

General IconGeneral
BrandSiemens
ModelSOMATOM
CategoryMedical Equipment
LanguageEnglish

Related product manuals