EasyManuals Logo
Home>Siemens>Medical Equipment>SOMATOM

Siemens SOMATOM Manual&Nbsp;

Siemens SOMATOM
590 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #49 background imageLoading...
Page #49 background image
Dose Information
49
How does CARE Dose 4D work?
CARE Dose 4D combines two types of tube current
modulation:
1) Axial tube current modulation:
Based on a single Topogram (a.p. or lateral) the atten-
uation profile along the patient’s long axis is measured
in direction of the projection and estimated for the per-
pendicular direction by a sophisticated algorithm.
Example of lateral and a.p. attenuation profile evalu-
ated from an a.p. Topogram.
Based on these attenuation profiles, axial tube current
profiles (lateral and a.p.) and the resulting eff. mAs for
every table position are calculated. The correlation
between attenuation and tube current is defined by an
analytical function which results in an optimum dose
and image noise in every slice of the scan.
Attenuation (log)
Lateral
Scan Range

Table of Contents

Other manuals for Siemens SOMATOM

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens SOMATOM and is the answer not in the manual?

Siemens SOMATOM Specifications

General IconGeneral
BrandSiemens
ModelSOMATOM
CategoryMedical Equipment
LanguageEnglish

Related product manuals