Safety Instructions
HEC002-A/HEC006-A 2-5
2.3.1 Interlock list
Table2-3 Safety interlock list (Part 1)
Overheat in
Heat
Exchanger
Detects abnormal heat in heat
exchanger.
Switching
power supply
failure
Detects abnormal current, voltage or
heat.
Detects overheat in Heat Sink.
Low circulating
fluid level
Detects lowering of circulating fluid level
in reservoir.
Detection of
breakage of
temp. sensor
Controller
internal circuit
Detects breakage and short circuit of
temp. sensor and cables.
Controller
internal circuit
Detects temp. input value not less than
70 C.
Controller
internal circuit
Detects temp. input value not more than
0 C.
Abnormal
output
(Overload)
Controller
internal circuit
Works when temp. change is smaller
than over load detection temp.(selected
between 0.1 to 9.9 C) even if 100%
output continues during overload
detection time (selected between 0 to
99min.) When 0 is set, the alarm doesn’t
arise.
Detects lowering of flow rate at
0.7±0.3L/min or less.
Table2-4 Safety interlock list (Part 2)
Overheat in Heat
Exchanger
Shuts off power supply to
the Heat exchanger, Pump
and Fan.
Error indication on display.
(ERR14)
Switching power
supply failure
Shuts off power supply to
the Heat exchanger, Pump
and Fan.
Error indication on display.
(ERR11)
Shuts off power supply to
the Heat exchanger, Pump
and Fan.
Error indication on display.
(ERR14)
Low circulating fluid
level
Shuts off power supply to
the Heat exchanger, Pump
and Fan.
Error indication on display.
(ERR20)