Operation
HEC002-A/HEC006-A
Table8-2 Available mode in Level 2
Setting range
(Min. increment)
Fine Control of
Internal Sensor
Sets the fine adjusting value to
calibrate the internal temp. sensor.
-9.99 to 9.99 C
(0.01 C)
Fine Control of
External Sensor
Sets the fine adjusting value to
calibrate the external temp. sensor
available optionally.
-9.99 to 9.99 C
(0.01 C)
Sets PB (Proportional Band) range used
for PID control.
Sets integral operation range of PID
control.
ARW: Anti Reset Windup
Sets integral time used for PID control.
Sets differential time used for PID
control.
When 0 is set, differential operation is
not made.
Sets output ratio of cooling to heating to
compensate difference of gain between
them.
Overload Judging
Temp. Range
Sets the temp. range for judgment of
overload(accompanying abnormal
output alarm ERR15).
Sets time for judgment of overload
(accompanying abnormal output alarm
ERR15). When 0 is set, the alarm
doesn’t arise.
Shows output ratio of thermo module
by 1%. The prefix symbol “-” stands for
cooling and no prefix stands for
heating.
Upper/Lower Temp.
Alarm Sequence
Determines whether or not temp.
upper/lower limit alarm is output when
power is turned on.
On : Output
Off : Not output
* How to return default value: Turn on the power supply with pressing [SEL] and [RET] keys. At this time, the
settings of Level 1 and Level 3 are also reset.
* The default value indicated in ( ) is an optimal value. The value is inputted at that time of shipment factory. If
turn on the power supply with pressing [SEL] and [RET] keys, set up the value indicated in ( ).