REV JUL 1987
Video Option characteristics are given in Table 1-4.
Mechanical characteristics of the 2430 are listed in
Table 1-3.
Environmental Characteristics are given in Table 1-2.
The 2430 meets the environmental requirements of MIL-
T-28800C for Type III, Class 3, Style D equipment, with
the humidity and temperature requirements defined in
paragraphs 3.9.2.2, 3.9.2.3, and 3.9.2.4. The rackmounted
2430 meets the vibration and shock requirements of MIL-
T-28800C for Type III, Class 5, Style D equipment when
mounted using the rackmount rear-support kit supplied
with both the 1R Option and the Rackmount Conversion
kit.
A complete adjustment procedure that includes the
EXTENDED CAL and all the internal 2430 adjustments
should normally be done after 2000 hours of operation or
at one-year intervals if used infrequently.
2. Immediately before making measurements requiring
the highest degree of accuracy.
1. If the operating temperature has changed by more
than 5°C since the last SELF CAL was done.
For optimum performance to specification, the 2430
SELF CAL may be done:
Items listed in the "Performance Requirements" column
are verifiable quantitative or qualitative limits that define
the measurement capabilities of the 2430 Oscilloscope.
The following electrical characteristics (Table 1-1) apply
when the 2430 has been calibrated at an ambient tem-
perature between +20°C and +30°C, has had a warm-
up period of at least 20 minutes and is operating at an
ambient temperature between -15°C and +55°C (unless
otherwise noted).
PERFORMANCE CONDITIONS
For part numbers and further information about stand-
ard accessories and a list of the optional accessories, refer
to "Options and Accessories" (Section 7) either in this
manual or in the Operators Manual. For additional informa-
tion on accessories and ordering assistance, contact your
Tektronix representative or local Tektronix Field Office.
'·2
2 Probe packages
Snap-lock accessories pouch
Zip-lock accessories pouch
Operators manual
Instrument Interface Guide
2 User Reference Guide
Fuse
Power cord (installed)
Blue plastic crt filter (installed)
Clear plastic crt filter
Front-panel cover
The following items are standard accessories shipped
with the 2430 instrument:
The 2430 has full TALKER/LISTENER GPIB capabilities
that make the instrument ideal for making automated
measurements in an installed system and a standard X-Y
Recorder output for producing low-cost hard copies of
acquired waveforms. The GPIB interface may also be used
to drive a ThinkJet® printer to plot the acquired
waveforms.
Time and Voltage cursors are provided for making
measurements on the displayed waveforms. Direct readout
of the measured values are displayed on the crt. The cur-
sors have alternate modes of readout operation for a
choice of measurement types (i.e., volts at time, slope, fre-
quency, absolute, dB, percent, degrees of phase, and delta
measurements).
Acquired waveforms may be saved in any of four REF
waveform nonvolatile memories. Any or all of the saved
reference waveforms may be displayed for comparison
with the waveforms being currently acquired. The source
and destination of waveforms to be saved may be user
designated. Up to five front-panel control setups may be
saved (when REF4 is designated for front-panel memory)
in nonvolatile memory for recall at a later time.
The ENVELOPE Mode display presents a visual image
of the amount of change (envelope) that occurs to a
waveshape during the acquisition of from 1 to 256
waveforms. The continuous ENVELOPE Mode holds all
changes in the display until reset by a control change.
ENVELOPE Mode can capture single-event pulses
(glitches) as narrow as 4 ns at the slowest SEC/DIV set-
ting of 5 seconds per division.
may be used in NORMAL and AVG Modes for equivalent-
time sampling that extends the useful storage bandwidth
to 150 MHz for recurring periodic signals.