EasyManuals Logo
Home>Tektronix>Test Equipment>2445

Tektronix 2445 User Manual

Tektronix 2445
55 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #15 background imageLoading...
Page #15 background image
Specification—2445 Operators
Table 1-1 (cont)
Characteristics
Performance Requirements
SIGNAL OUTPUTS (cont)
CALIBRATOR (cont)
Symmetry Duration of high portion of output cycle is 50% of output period
± (lesser of 500 ns or 25% of period).
Jitter of Pulse Period or Pulse Width 10 ns or less.
CH 2 SIGNAL OUT
Output Voltage
20 mV per division ±10% into 1 M£2;
10 mV per division ±10% into 50 £2.
Offset
±10 mV into 50 f2, when dc balance has been performed within
±5°C of the operating temperature.
A GATE OUT and B GATE OUT
Output Voltage
2.4 V to 5 V positive-going pulse, starting at 0 V to 0.4 V.
Output Drive
Will supply 400 juA during HI state; will sink 2 mA during LO state.
CRT
Display
80 mm x 100 mm.
Standard Phosphor
P31.
Nominal Accelerating Potential
16 kV.
AC POWER SOURCE
Source Voltage
Ranges
115 V
90 V to 132 V.
230 V
180 V to 250 V.
Source Frequency
48 Hz to 440 Hz.
Fuse Rating
2 A, 250 V, AGC/3AG, Fast blow;
or 1.6 A, 250 V, 5 x 20 mm, Quick-acting (F).
Power Consumption
Typical
70 W (140 VA).
Maximum
120 W (180 VA).
Primary Circuit Dielectric Voltage Withstand Test 1500 V rms, 60 Hz for 10 s without breakdown.
Primary Grounding
Type test to 0.1 f2 maximum. Routine test to check grounding
continuity between chassis ground and protective earth ground.
1-8