Display Waveform Data
7. Select Fine Scale to enable multipurpose
knob a to make ne vertical scale adjustments.
Fine Scale
8. Select Offset to enable multipurpose knob a
to make vertical offset adjustments.
On the side-bezel menu, choo se Set to 0 V to
set the vertical offset to 0 V.
For m o re information on offset, see Quick
Tips. (See page 161, Quick Tips.)
Offset
Position
Deskew
9. Select Deskew to set the time skew correction
for the channel. Turn multipurpose knob a to
adjust the time skew (deskew) correction for
the probe attached to the selected channel.
This shifts acquisition and display of the
waveform left or right, relative to the trigger
time. Use this to compensate for differences in
cable lengths or probe types.
160 DPO2000 and MSO2000 S eries Oscilloscopes User Manual