Acquire the Sign
al
To acquire data from an I
2
C bus, you need to also set up these items:
1. If you select I
2C, push Dene Inputs and
the desired side-bezel menu choices.
B1
I2C
Dene
Inputs
Thresholds Include
R/W in
Address
No
B1 Label
I2C
Bus
Display
Event
Table
You can assign the prede ned SCLK Input
or SD A Input to the channel connected to
the signal.
2. Push Include R/W in Address and then
push the des
ired side-bezel button.
This control determines how the oscilloscope
shows the I
2
C addresses in bus decode
traces, cur
sor readouts, Event Table listings,
and trigger settings.
If you sele
ct Yes, the oscilloscope displays 7-bit addresses as eight bits, where the eighth bit (LSB) is the R/~W bit.
If you sel
ect No, the oscilloscope displays 7-bit addresses as seven bits, and 10-bit addresses as ten bits.
The oscil
loscope also displays 10-bit addresses as 11 bits. The rst two bits are the two MSBs of the address. The next
bit is the R/~W bit. The last eight bits are the eight LSBs of the address. (In the physical layer of the I
2
C protocol, 10 bit I
2
C
addresses are preceded by the ve bit code, 11110. The oscilloscope never includes these ve bits in address readouts.)
54 MSO2000B and DPO2000B Series Oscilloscopes User Manual

Table of Contents

Other manuals for Tektronix DPO2012B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO2012B and is the answer not in the manual?

Tektronix DPO2012B Specifications

General IconGeneral
BrandTektronix
ModelDPO2012B
CategoryTest Equipment
LanguageEnglish

Related product manuals