Trigger Setup
If you have made a Trigger On selection
of Data or Address/Data, push the
lower-bezel D
ata button to access the I
2
C
Data side-bezel menu.
Push the Number of Bytes button and
enter the num
ber of bytes with multipurpose
knob a.
Push the side-bezel Addressing Mode
button and se
lect 7bitor 10 bit. Push
the side-bezel Data button. Enter the data
parameters of interest with multipurpose
knobs a and b
.
For more information on the I
2
C address
formats, refer to item 2 under Setting Up
Bus Parame
ters.
8. If you are setting up an SPI trigger and have
made a Tri
gger On selection of MOSI or
MISO, push the lower-bezel Data button,
push the side-bezel MOSI (or MISO) button,
and enter
the data parameters of interest
with multipurpose knobs a and b.
Then push the Number of Bytes button and
enter th
e number of bytes with multipurpose
knob a.
If you select MO SI & MISO, push the
lower-
bezel Data button and enter the
parameters of interest in the side-bezel
menus.
9. If you are setting up a CAN trigger and have
made a Trigger On selection of Type of
Frame
, push the lower-bezel Frame Type
button, and select Data Frame, Remote
Frame, Error F rame, or Overload Frame.
If yo
u have made a Trigger On selection of
Identifier, push the lower-bezel Identifier
button, and select a Format. Then push the
Iden
tifier side-bezel button, and enter a
binary or hex value with multipurpose knobs
a and b.
Pus
h the lower-bezel menu Direction button
and select the direction of interest: Read,
Write,orRead or Write.
If y
ou have made a Trigger On selection
of Data. Push the lower-bezel Data button
and enter the parameters of interest w ith
mu
ltipurpose knobs a and b.
72 MSO2000B and DPO2000B Series Oscilloscopes User Manual