2260B Series User Manual
124
Connected to the (–S) sense- terminal when
remote sense is used. Connected to the negative
output terminal when remote sense is not used.
Current sum output signal when used in parallel
mode.
External voltage control of the voltage output. A
voltage of 0~10V is used to control the full scale
voltage output (0%~100%) of the instrument.
External voltage control of the current output. A
voltage of 0~10V is used to control the full scale
current output (0%~100%)of the instrument
External resistance control of the voltage output. A
resistance of 0kΩ ~ 10kΩ is used to control the full
scale voltage output (0%~100%) of the instrument.
External resistance control of the voltage output. A
resistance of 0kΩ ~ 10kΩ is used to control the full
scale voltage output (0%~100%) of the instrument.
External resistance control of the current output. A
resistance of 0kΩ ~ 10kΩ is used to control the full
scale current output (0%~100%) of the instrument.
External resistance control of the current output. A
resistance of 0kΩ ~ 10kΩ is used to control the full
scale current output (0%~100%) of the instrument.
Voltage Monitor Output. Outputs the full scale
voltage (0~100%) as a voltage (0V~10V).
Current Monitor Output. Outputs the full scale
current (0~100%) as a voltage (0V~10V).
The shut down signal will turn off the output or
power when a low TTL signal is applied. The
shutdown signal is pulled up to 5V with a 10kΩ
pull-up resistor.
Master unit current sum input signal from first
slave CURRENT SUM OUTPUT. Used in parallel
mode only.
Master unit current sum input signal from second
slave CURRENT SUM OUTPUT. Used in parallel
mode only.