2260B Series User Manual
134
4. Press the Output key. The current
can now be controlled with the
External resistance.
Note
Ensure the resistor(s) and cables used exceed the
isolation voltage of the power supply. For example:
insulation tubes with a withstand voltage higher
than the power supply can be used.
When choosing an external resistor ensure the
resistor can withstand a high degree of heat.
Note
CV and CC Slew Rate Priority are disabled for V-I
mode (F-03) when using external resistance
control. See the normal function settings on page
108.
External Control of Output
Background
The output can be turned on or off externally
using a switch. The analog control connector
can be set to turn the output on from a high or
low signal. The voltage across pins 2 and 24 are
internally pulled to +5V ±5% @ 500uA with
10kΩ pull-up resistor. A short (closed switch)
produces a low signal.
When set to High = On, the output is turned on
when the pins 2-24 are open.
When Low = On, the output is turned on when
pins 2-24 are shorted.

Table of Contents

Other manuals for Tektronix Keithley 2260B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley 2260B Series and is the answer not in the manual?

Tektronix Keithley 2260B Series Specifications

General IconGeneral
BrandTektronix
ModelKeithley 2260B Series
CategoryPower Supply
LanguageEnglish

Related product manuals