ANALOG CONTROL
123
Analog Remote Control Overview
The 2260B power supply series have a number of analog control
options. The Analog Control connectors are used to control output
voltage and current using external voltage or resistance. The power
supply output and power switch can also be controlled using
external switches.
Analog Control connector overview → from page 123
External voltage control of voltage output from page 125
External voltage control of current output → from page 128
External resistance control of voltage output → from page 130
External resistance control of current output → from page 132
External control of output → from page 134
External control of the power switch → from page 137
Analog Control Connector Overview
Overview
The Analog Control Connector is a standard
Mil 26 pin connector (OMRON XG4 IDC plug).
The connector is used for all analog remote
control. The pins used determine what remote
control mode is used.
WARNING
To prevent electric shock, ensure that the cover for
the Analog Control Connector is used when the
connector is not in use.
Pin Assignment
1
26
25
2
Pin name
Pin number
Description
Current Share
1
Used when operating 2 or more units in parallel.

Table of Contents

Other manuals for Tektronix Keithley 2260B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley 2260B Series and is the answer not in the manual?

Tektronix Keithley 2260B Series Specifications

General IconGeneral
BrandTektronix
ModelKeithley 2260B Series
CategoryPower Supply
LanguageEnglish

Related product manuals