OPERATION
63
Set to CV Mode
When setting the power supply to constant voltage mode, a current
limit must also be set to determine the crossover point. When the
current exceeds the crossover point, the mode switches to CC mode.
For details about CV operation, see page 25. CC and CV mode have
two selectable slew rates: High Speed Priority and Slew Rate
Priority. High Speed Priority will use the fastest slew rate for the
instrument while Slew Rate Priority will use a user-configured slew
rate.
Before setting the power supply to CV mode,
ensure:
The output is off.
The load is connected.
1. Press the Function key. The
Function key will light up.
2. The display should show F-
01 on the top and the
configuration setting for
F-01 on the bottom.
3. Rotate the Voltage knob to change
the F setting to F-03 (V-I Mode
Slew Rate Select).
4. Use the Current knob to set the F-
03 setting.
Set F-03 to 0 (CV High Speed
Priority) or 2 (CV Slew Rate
Priority).
0 = CV High Speed Priority
2 = CV Slew Rate Priority