OPERATION
61
ALM
V
A
Before setting the OVP or OCP level:
Ensure the load is not connected.
Ensure the output is set to off.
Setting Ranges
2260B (360W)
30-36
80-13
250-4
800-1
OVP Range (V)
3-33
8-88
20-275
20-880
OCP Range (A)
3.6-39.6
1.35-14.85
0.45-4.95
0.144-1.584
2260B (720W)
30-72
80-27
250-9
800-2
OVP Range (V)
3-33
8-88
20-275
20-880
OCP Range (A)
5-79.2
2.7-29.7
0.9-9.9
0.288-3.168
2260B (1080W)
30-108
80-40
250-13
800-4
OVP Range (V)
3-33
8-88
20-275
20-880
OCP Range (A)
5-118.8
4.05-44.55
1.35-14.85
0.432-4.752
Steps
1. Press the OVP/OCP key. The
OVP/OCP key lights up.
2. The OVP setting will be displayed on the top
and the OCP setting (or OFF) will be displayed
on the bottom.
V
A
OVP Setting
OCP Setting

Table of Contents

Other manuals for Tektronix Keithley 2260B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley 2260B Series and is the answer not in the manual?

Tektronix Keithley 2260B Series Specifications

General IconGeneral
BrandTektronix
ModelKeithley 2260B Series
CategoryPower Supply
LanguageEnglish

Related product manuals