Operating Requirements
Figure 1: MDO3000 series
TPP0250, TPP0500B, or
TPP1000 Pass ive Probe
Maximum input voltage: 300 V
RMS
CAT II safety requirements
Temperature:
Operating:1Cto+6C(+Fto+14F)
Nonoperating: –62 °C to +85 °C ( –80 °F to +185 °F)
Humidity: 5% to 95% RH
Operating: 5% to 95% relative humidity (%RH) up to +30 °C, 5% to 75% RH
above +30 C up to +65 °C. Noncondensing
Nonoperating: 5% to 45% RH above +65° C up to +85 °C. Noncondensing
Altitude:
Operating: 3,000 m (9,842 ft.) maximum
Nonoperating: 12,200 m (40,000 ft.) maximum
Option MDO3MSO: Adds
16 digital channels and a
P6316 Digital Probe
Number of input channels: 16 digital inputs
Threshold Accuracy: ±(100 mV + 3% of threshold)
Threshold Range: –15 V to +25 V
Maximum nondestructive input signal to probe: –20 V to +30 V peak
Minimum signal swing: 500 mV
peak-to-peak
Input resistance: 101 k
Input capacitance: 8.0 pF
Temperature:
Operating: 0 °C to +50 °C (+32 °F to +122 °F)
MDO3000 Installation and Safety Instructions 15

Table of Contents

Other manuals for Tektronix MDO3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO3000 Series and is the answer not in the manual?

Tektronix MDO3000 Series Specifications

General IconGeneral
BrandTektronix
ModelMDO3000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals