Installation
2. Check the shape of the displayed
waveform to determine if your
probe is prope
rly compensated.
Properly compensated
Under compensa
ted
Over compensat
ed
3. If necessary,
adjust your probe.
Repeat as needed.
Quick Tips
Use the shortest possible ground lead
and signal path to minimize probe-induced
ringing a
nd distortion on the measured
signal.
Signal with a short ground lead
Signal w
ithalonggroundlead
Application Module Free Trial
A 30-day free trial is available for all application module licenses not installed in y our oscilloscope. The trial period begins
when you power on the oscilloscope for the rst time.
After 30 days, you must purchase the module if you want to continue using the application. To see the date when your
free trial period expires, push Utility on the front panel, push Utility Page on the lower menu, turn the Multipurpo se a
knob to select Cong, and push About on the lower menu.
Installing an Application Module
CAUTION. To avoid damage to the oscilloscope or application module, observe ESD (electrostatic discharge) precautions.
(See page 9, Powering on the Oscilloscope.)
Turn off the oscilloscope power while removing or adding an application module.
(See page 10, Powering off the Oscilloscope.)
14 MDO3000 Series Oscilloscopes User Manual

Table of Contents

Other manuals for Tektronix MDO3054

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO3054 and is the answer not in the manual?

Tektronix MDO3054 Specifications

General IconGeneral
BrandTektronix
ModelMDO3054
CategoryTest Equipment
LanguageEnglish

Related product manuals