Specications
Environmental performance
Temperature
Operating –10 °C to +55 °C (+14 °F to +131 °F)
Nonoperatin
g
–51 °C to +71 °C (–60 °F to +160 °F)
Humidity, o p
erating
5% to 95 ±5% RH (relative humidity) in the temperature range of +10 °C to 30 °C
(+5Fto8F)
5% to 75% ±5%
RH from +30 °C to +40 °C (+86 °F to 104 °F)
5% to 45% RH above +40 °C to +55 °C (+86 °F to +131 °F)
<10 °C humidity is uncontrolled; non-condensing
Altitude
Operating 9144 meters (30,000 feet)
Nonopera
ting
15,240 meters (50,000 feet)
Dynamic
s
Random vibration:
nonoper
ating
0.030 g
2
/Hz, 10 Hz 500 Hz, 30 minutes per axis, three axes (90 minutes total)
Mechanical shock: operating
Half-s
ine mechanical shocks, 30 g peak amplitude, 11 ms duration, three drops
in each direction of each axis (18 total)
Handli
ng and transit
Bench handling, operating
Per MI
L-PRF-28800F Class 2 operating: Rotational-edge-drops of appropriate edges
on appropriate sides of the equipment
Trans
it drop, nonoperating
Per MIL-PRF-28800F Class 2 nonoperating: Transit drops onto six faces and four
corners of the equipment, from a height of 30 cm (11.8 in.) for a total of 10 impacts
RSA306B Specications and Performance Verication 11

Other manuals for Tektronix RSA306B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix RSA306B and is the answer not in the manual?

Tektronix RSA306B Specifications

General IconGeneral
BrandTektronix
ModelRSA306B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals