Specications
Frequency
RF input frequency range 9kHzto6.2GHz
Frequency reference accuracy
Over 18 °C to 28 °C ambient
temperature range
±3 x 10
–6
+ aging (after 20 minute warm up)
Aging, typical
±3 x 10
–6
(rst year), ±1 x 10
–6
each year thereafter
Over operating temperature
range (–10
°C to +55 °C
ambient), typical
±20 x 10
–6
+ aging
External frequency reference
input
Input Frequency Range 10 MHz ± 10 Hz
Input Lev
el Range
–10 dBm to
+10 dBm sinusoid
Impedance
50
Center
frequency resolution
Block I
Q samples
1Hz
Strea
med ADC samples
500 kHz
Amplitude
RF input impedance
50
RF input VSWR, typical 1.8:1 (10 MHz to 6200 MHz, reference level +10 dBm)
(Eq
uivalent Return Loss: 11 dB)
Maximum RF input level without
dam
age
The maximum voltage or power that the RF input can withstand without creating
ash
ock hazard or damaging the input.
DC
voltage
±40 V
DC
Re
fLevel –10 dBm
+2
3 dBm (continuous or peak)
R
ef Level < –10 dBm
+
15 dBm (continuous or peak)
Maximum RF input operating
l
evel
T
he maximum level at the RF input for which the instrument will meet its
measurement specications.
Center frequency < 22 MHz
+15 dBm
Center frequency 22 MHz
+20 dBm
2 RSA306B Specications and Performance Verication

Other manuals for Tektronix RSA306B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix RSA306B and is the answer not in the manual?

Tektronix RSA306B Specifications

General IconGeneral
BrandTektronix
ModelRSA306B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals