Maintenance
Keypad
You will need a t
orque-limiting Torx T-15 screwdriver for this procedure.
Refer to the exploded view diagram for the 2-channel models or the 4-channel
models. (See p
age 8-3.) (See Figure 8-2.)
Removal.
1. Remove the p
ower button and rear case. (See page 6-10, Rear Case.)
2. Remove the internal a ssembly from t he front case. (See pag e 6-13, Internal
Assembly.
)
3. Lift the keypad switch out of the front case.
CAUTION.
Keep the electrical contacts on the back of the keypad clean. Skin oils
and dust inhibit good electrical contact.
Install
ation.
1. Place the front case face down on a soft su rface (such as an anti-static mat).
2. Align the switch keypad into the front case using the guide posts and the
holes in the keypad.
CAUTION. Keep the electrical contacts on the back of the switch keypad clean.
Skin
oils and dust inhibit good electrical contact.
3. Check that all b uttons are properly seated before proceeding.
4. Assemble the oscilloscope by p erforming the installation procedure in reverse
order for each module that was removed.
TBS1000 Series Oscilloscope Service Manual 6–21

Table of Contents

Other manuals for Tektronix TBS1000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TBS1000 Series and is the answer not in the manual?

Tektronix TBS1000 Series Specifications

General IconGeneral
BrandTektronix
ModelTBS1000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals