Brief Procedures
18
TDS 500D, TDS 600C, TDS 700D & TDS 714L Performance Verification and Specifications
Equipment
required
One probe such as the P6243, P6245, P6249, P6139A
1
, or P6339A
Prerequisites None
1
P6139A probe is not appropriate for the TDS694C/794D oscilloscopes.
1. Install the test hookup and preset the oscilloscope controls:
a. Hook up the signal source: Install the probe on CH 1. Connect the probe
tip to PROBE COMPENSATION SIGNAL on the front panel;
connect the probe ground (typically black) to PROBE COMPENSA-
TION GND. If using a P6243, P6245 or P6249 probe, you may want to
attach a Y-lead connector and two SMD KlipChips as shown in
Figure 13.
Signal Gnd
Figure 13: Universal test hookup for functional tests TDS 600C shown
b. Initialize the oscilloscope:
H Press save/recall SETUP.
H Press the main-menu button Recall Factory Setup.
H Press the side-menu button OK Confirm Factory Init.
Verify All Input Channels
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Tektronix TDS 520D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 520D and is the answer not in the manual?

Tektronix TDS 520D Specifications

General IconGeneral
BrandTektronix
ModelTDS 520D
CategoryTest Equipment
LanguageEnglish

Related product manuals