Performance Tests
146
TDS 500D, TDS 600C, TDS 700D & TDS 714L Performance Verification and Specifications
TRIGGER LEVEL knob to do this. (Full bandwidth varies with
TDS model as is shown in Table 14).
g. Measure the test signal:
H Set the frequency of the generator, as shown on screen, to the test
frequency in Table 14 that corresponds to the vertical scale set in
substep d. See Figure 19.
H Set the horizontal SCALE to the horizontal scale setting in
Table 14 that corresponds to the vertical scale set in substep d.
Press SET LEVEL TO 50% as necessary to trigger the signal.
H Read the results at the CHx Pk-Pk readout, which will automatically
measure the amplitude of the test signal. See Figure 19.
Set the generator
(reference) frequency to
the test frequency from
Table 14.
1
Set the horizontal scale
from Table 14.
2
Read results.
3
Figure 19: Measurement of analog bandwidth
h. Check against limits:
H CHECK that the Pk-Pk readout on screen is within the limits listed
in Table 14 for the current vertical scale setting.
H Enter voltage on test record.
H When finished checking, set the horizontal SCALE back to the
50 ns (TDS694C 40 ns) setting.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Tektronix TDS 520D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 520D and is the answer not in the manual?

Tektronix TDS 520D Specifications

General IconGeneral
BrandTektronix
ModelTDS 520D
CategoryTest Equipment
LanguageEnglish

Related product manuals