Adjustment Procedures
TDS1000B and TDS2000B Series Oscilloscope Service Manual
5-- 3
7. Release the SINGLE SEQ button.
8. Release the AUTOSET button. A message appears in the lower left corner of
the screen stating “Service mode ON.”
9. Push the front-panel UTILITY button. The last item in the Utility menu is
now “Service.”
At completion of the Adjust procedure disable the “Service” menu through the
UTILITY front panel button, the “Service” option button, and the “Service”
Mode Off” option button.