DTM500 Series Specifications
2
Handheld and Benchtop Instruments Basic Service
Specifications
The characteristics listed in this section apply under the following conditions:
H The instrument operates in an 18° to 28° C ambient environment unless
otherwise noted.
H The instrument warms up for 60 minutes.
NOTE. All specifications are warranted unless marked “typical.” Typical
characteristics are not guaranteed but are provided for the convenience of the
user.
Specifications marked with the n symbol are checked in the performance
verification procedures.
Table 1: General Characteristics
Characteristic Description
Product Measurement Range
DTM510 –50_ to 1300_ C (–58_ to 1999_ F)
DTM520 (K-type probe) –200_ to 1370_ C (–328_ to 2498_ F)
DTM520 (J-type probe) –200_ to 760_ C (–328_ to 1400_ F)
nAccuracy
Temperature Range Resolution % of Reading
DTM510 –50_ to 199.9_ C 0.1_ C ±(0.2% + 1_ C)
–58_ to 391_ F 0.1_ F ±(0.2% + 2_ F)
–50_ to 199.9_ C 1_ C ±(0.3% + 2_ C)
–58_ to 391_ F 1_ F ±(0.3% + 4_ F)
200_ to 999.9_ C 1_ C ±(0.4% + 2_ C)
392_ to 1831_ F 1_ F ±(0.4% + 4_ F)
1000_ to 1300_ C 1_ C ±(0.6% + 2_ C)
1832_ to 1999_ F 1_ F ±(0.6% + 4_ F)