TAS 200 Series Adjustment Procedures
22
Handheld and Benchtop Instruments Basic Service
Test Equipment
To ensure accurate adjustments, use the following or equivalent test equipment.
If you substitute equipment, always choose instruments that meet or exceed the
minimum requirements specified in Table 19.
Alternative test equipment must meet or exceed the intended minimum
requirements. If you substitute equipment, you may need to modify the
adjustment procedures.
NOTE. Before making any adjustment, warm up the test equipment according to
the manufacturer’s recommendations.
Table 19: Adjustment Test Equipment
Description Minimum Requirements Example Product
Leveled Sine Wave Generator 1 kHz to 250 MHz; variable amplitude from
5 mV to 4 V
p-p
into 50 W
Wavetek
1
Universal
alibration
System with Oscilloscope Calibration
Time Mark Generator Variable marker frequency from 1 ms to
10 ns; accuracy within 2 ppm
Module (Option 2
)
Fluke 5500A Multi-product Calibrator with
DC Calibration Generator DC voltage levels from 100 mV to 10 V
Oscilloscope Calibration Option 5500A-SC
Digital Multimeter 35 V, 4 A, ±0.1% Accuracy Tektronix DMM252 Digital Multimeter
Termination (two required) 50 W, female BNC input, male BNC output Tektronix 011-0049-01
Cable, Coaxial 75 W, male-to-male BNC connectors,
36 inch length
Tektronix 012-1338-00
Cable, Precision Coaxial 50 W, male-to-male BNC connectors,
36 inch length
Tektronix 012-0482-00
Coupler Female-BNC-to-dual-male-BNC, dual-input Tektronix 067-0525-02
Probe Standard accessory probe Tektronix P6109B