25
4
Troubleshooting
Table
4
lists the symptoms, possible causes, and recommended
corrective actions for problems you may encounter with the
Certifier
®
FA Test System. If
the symptom is not listed, or if none of
the recommended corrective actions solves the problem, please
contact TSI Customer Support at (800) 874-2811 or 651-
490
-2811.
Table 4. Troubleshooting the Certifier
®
FA Test Sy
st
em
Symptom
Possible Cause
Corrective Action
Controller module
won’t turn on, or
turns on and off.
Batteries are
depleted or installed
backwards.
Check that batteries
are installed correctly.
Replace batteries.
Measurements
aren’t displayed
even though
controller module
is on.
Flow module isn’t
connected to
controller module.
Connect flow module to
controller module.
‘
---
‘ is shown on
display.
Measurement is out
of range.
Check range for
displayed
measurement, and only
make measurements
within that range.
Flow rate does
not read zero
when no gas
flowing
(Especially when
set to N
2
O).
Meter was not
purged with gas
displayed on the
controller.
Purge meter with gas
displayed on controller
or press gas select key
to change to desired
gas.
Unable to
disconnect flow
module from
controller module.
Pulling on the cable
rather than the
connector.
Pull the locking
connector (not the
cable) to disengage
connector lock.
Can’t display
%O
2
.
Oxygen sensor not
connected. Look for
oxygen sensor
symbol,
.
Connect oxygen sensor
cable to High Flow
module and oxygen
sensor.
32
34
Table of Contents
Default Chapter
7
Table of Contents
7
1 Introduction
9
Figure 1. the Certifier ® FA Test System
10
Parts List
11
Figure 2. Certifier FA Test System Parts
11
Table 1. Certifier FA Test System Parts List
12
Glossary
14
2 Setup
15
Figure 3. Connecting the Controller Module to a Flow Module
15
Figure 4. Connecting the Bacteria Filters to the Flow Modules
16
Figure 5. Attaching Pressure Tubing to the High Flow Module
17
Figure 6. Attaching the Oxygen Sensor to the High Flow Module
18
Figure 7. Installing a Flow Module into the Test Circuit
19
Figure 8. Certifier FA Test System Installed in an Oxygen Concentrator Circuit
20
3 Operation
21
Power up
21
Keypad Functions
22
Displaying Test Measurements
22
Figure 9. Controller Module Keypad
22
Table 2. Keypad Functions and Operation
22
Measurements
23
Flow Rate
23
Peak Flow Rate
24
Volume
24
Minute Volume
24
Stacked Volume
25
Oxygen Concentration
25
Low Pressure
25
Peak Pressure
26
Peep Pressure
26
Breath Rate
26
I:E Ratio
26
I Time
27
Absolute Pressure
27
Oxygen Concentration
27
Required Pre-Test Calibrations
28
Low-Pressure Transducer Zero Calibration
28
Oxygen Sensor Calibration
28
Breathing Cycles and Trigger Levels
29
Inhalation and Exhalation Timing
29
Flow Trigger Levels
29
Display Information
30
Table 3. Screen Displays
30
Troubleshooting
33
Table 4. Troubleshooting the Certifier
33
Maintenance
37
Replacing the Batteries (as Required)
37
Replacing the Oxygen Sensor (Yearly)
37
Cleaning (as Required)
37
Table 5. Cleaning Recommendations
37
Factory Calibration (Yearly)
38
Return Procedure
38
6 Specifications
39
Physical
39
Environmental
39
Power
39
Test Measurements
39
Calibration Requirements
41
Compliance and Approvals
41
Module
41
5
Based on 1 rating
Ask a question
Give review
Questions and Answers:
Need help?
Do you have a question about the TSI Instruments CERTIFIER FA and is the answer not in the manual?
Ask a question
TSI Instruments CERTIFIER FA Specifications
General
Brand
TSI Instruments
Model
CERTIFIER FA
Category
Test Equipment
Language
English