0 to 150.0 cmH
2
O.
(0 to 110 mmHg)
0.75% of reading or 0.20
cmH
2
O (0.15 mmHg), whichever
is greater.
2% O
2
at daily calibration
conditions.
NOTES
1. Standard conditions are defined as 21.1 C (70 F) and 101.3 kPa (14.7 psia).
2. Flow and volume accuracy is applicable in SLPM or STP mode only.
3. The temperature of the gas and the ambient air must be within 10 C ( 18 F) of
each other and the gas must be less than 30% relative humidity at 21. 1 C (70 F).
4. Flow and volume accuracy de-rating: 0.075% of reading per 1 C (1.8 F) away
from 21.1 C (70 F); 0.015% of reading per 1.03 kPa (0.15 psia) above 101.3 kPa
(14.7 psia); 0.022% of reading per 1.03 kPa (0.15 psia) below 101.3 kPa (14.7
psia); 0.07% of reading per 1% relative humidity above 30% relative humidity.
6.5 Calibration Requirements
Factory calibration every year for normal use under normal
conditions using the filter(s) provided.
Daily, following sensor replacement or as required.
6.6 Compliance and Approvals
Complies with these
standards:
CAN/CSA-C22.2 No. 1010. 1-92: Canadian Standard for the
Safety of Electrical Equipment for Measurement, Control
and Laboratory Use, Part 1.
EN 55011 (1991) Class B, CISPR 11 (1990) Class B, FCC
(CFR 47, Part 15) Class B: Emissions, Radiated and
Conducted.
EMC Directive 89/336/ECC, EN 61326-1 (1997 plus
Amendment A1 1998), IEC 1000-4-2 (1995), EN 61000-4-2,
IEC 1000-4-3 (1995), EN 61000-4-3: Immunity.