Manual trigger: the instrument outputs pulse with over 1us pulse width from the beginning of
pulse train.
External trigger: the menu will hide the trigger output option which is enabled by external
modulation connector (FSK Trig connector). This connector should not be used as external and
internal at the same time.
3.3.9 Trigger Edge
Trigger edge can be designated when external digital modulation interface (FSK Trig connector) is
used as input. When it is used as input (i.e. internal trigger source), rising edgemeans that rising
edge of external signal triggers output of a pulse train, and falling edgemeans that falling edge of
external signal triggers output of a pulse train. In gating mode, when polarity in parameter list is
positive”, external signal triggers output of a pulse train at high level, and negative polarity
means that external signal triggers output of a pulse train at low level. When it is used as output (i.e.
internalor manualtrigger and trigger output is ON”), the default edge is rising edge”. If you
need to change, press
Param TrigEdge Fall (Gating mode: press Param Polarity
Negative )
Comprehensive Example
First make the instrument run in pulse train mode, and then set a sine wave signal with
period of 5ms and amplitude of 500mVpp as waveform of pulse train, set type of pulse train to be
N-cycle, period of pulse train to be 15ms and recurring number to be 2.The specific steps are as
follows:
Use pulse train function
Press Menu Burst Type
NCyc successively (press soft key Type to select if Type is not
highlighted) to set type of pulse train to be N-cycle mode.
Figure 3 - 52 Set N cycle function
Select waveform of Burst
After setting N-cycle mode of pulse train, press Carrier Type
Sine to select carrier signal as
sine wave. The default waveform of pulse train is sine wave, so it is unnecessary to change in this
example.