36
6Setupwiththedisplayandadjustmentmodule
MINITRAC 31 • Foundation Fieldbus
41782-EN-211203
Menu Menu item Default value
Setup Measurement loop name Sensor
Isotope Cs-137
Application Level
Background radiation 0ct/s
Process value unit %
Temperature unit ° C
Min. adjustment Min. process value = 0 %
Max. adjustment Max. process value = 100 %
Linearisation 0ct/s=100%
90000ct/s=0%
Damping 60 s
Real value correction 0
Mode Overllprotection
Upperswitchingpoint-Processvalue 0 %
Lowerswitchingpoint-Processvalue 0 %
Upperswitchingpoint-Temperature 50 °C
Upperswitchingpoint-Temperature 25 °C
Reference value - Relay None
Lock adjustment Released
Address - Summation Secondary free
Display Language Selected language
Displayed value Pulse rate
Display unit ct/s
Additional adjustments Temperature unit °C
Linearisation curve Empty
HART mode Standard
Address 0
6.4 Parameter adjustment - Density
measurement
The instrument is adapted to the application conditions via the pa-
rameteradjustment.Theparameteradjustmentiscarriedoutwithan
adjustment menu.
Prerequisites
Thefollowingrequirementsmustbefullledforreliableandsafe
operation:
•
Thetubemustbelled.Theremustbenoairbubblesinthetube
•
Thesourcecontainerisswitchedon
•
A sample point is close to the measuring point