HandyLab 100 Operation
41
ba77062d02 12/2017
Note
After the fifth buffer has been measured the calibration is automatically
finished. Then the calibration record is displayed.
The calibration line is determined by linear regression.
4.4.9 Displaying calibration records
Displaying the
calibration record
The calibration record of the last calibration is to be found under the
menu item, Calibration / Calibration record. To open it in the measured
value display, press the <CAL
_> key.
24 Set the nominal buffer value for the measured temperature with
<><>.
25 Accept the calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage dis-
play).
26 Use <M> to finish calibration or
switch to calibration with the next buffer with <ENTER>.
pH
9.958
24.8 °C
VariCal
3
15.03.2014
08:00
Buffer