<7. HART-COMMUNICATION>
7- 7
IM 01R01B02-00E-E 12th edition October 01, 2014 -00
All Rights Reserved. Copyright © 2003, Rota Yokogawa
7.4.1 Process variables
Device setup
Process variables
Flow value [cum/h] Flow in oriented measurement unit
Flow % rnge [%] %-value related to 20mA
Flow AO value [mA] Current output in mA
Totalizer value [kg] Totalizer
% span value %-Value related to ow nal value
Temperature value [°C] Temperature in transmitter
7.4.2 Diagnostic- and Service-Menu
Device setup
Diag/Service
Test device
Device status
Status group 1
Error indication:
RAM Error OFF/ON Memory error
ADC Error OFF/ON error A/D converter
ADJ-EE-Error OFF/ON error adjustment-EEPROM
CAL-EE Error OFF/ON error calibration-EEPROM
Totalizer false OFF/ON totalizer value false
Flow Overrun OFF/ON Flow too high
ADJ Error 2 OFF/ON adjustment-EEPROM fault
Float blocked OFF/ON Float-Move-Detection recognized, supervision
time passed
Device setup
Diag/Service
Test device
Device Status
Status group 2
Error indication:
Temp over limit OFF/ON Temperature exceed 70°C
Max ow1 active OFF/ON URV overrun
Max ow2 active OFF/ON URV underrun
FB autoz active OFF/ON Float blocking autozero on
Power fail warn OFF/ON Power off has happened
Oper timer error OFF/ON Operation timer fault
Mancal active OFF/ON Manual calibration activated