8-1
IM 760301-01E
8
Storing and Recalling Data and Saving the Stored Data
8.1 Setting the Store Mode
Procedure
START STOP
REMOTE
CAL
MOTOR
SET
CURSOR NULL
MENU STORE SET
MENU
LOWER ITEM LOWER FORM ALL
USER SET
DISPLAY
ITEM & ELEMENT
RESET SET
PAGE PAGE
SCALING
HRM SET
MEASURE
SYNC SOURCE
AVG
LINE FILTER
FREQ
FILTER
FILE
PRINT
MISC
STORE
IMAGE SAVE
HELP
NUMERIC
WAVE
OTHERS
ITEM
FORM
U / I / P
WP/q/
TIME
USER
ELEMENT
UPDATE
RATE
HOLD
SINGLE
INTEG
LOCAL
SHIFT
To exit the menu during operation, press ESC located above
the soft keys.
In the procedural explanation below, the phrase “press the
cursor keys” may be used. This phrase refers to the
procedures for selecting items and entering values and
character strings. For details on the procedures, see section
3.14.
Cursor keys
To set or execute the storage operation, set the measurement mode to normal
measurement mode or wide bandwidth harmonic measurement mode.* There are
limitations on the setting and execution of the storage operation in measurement
modes other than normal measurement mode.
* Can be set on models with the advanced computation (/G6) option.
1. Press SHIFT+STORE (STORE SET) to display the Store Set menu.
2. Press the Mode soft key to select Store.
3. Press the Store Setting soft key to display the Store Setting menu.
4. Press the Mode soft key to display the Mode menu.
5. Press any soft key from Manual to Integ Sync to select the store mode.
Chapter 8 Storing and Recalling Data and Saving the Stored Data