52
Understanding Your Control Results
Both the Accu-Chek Compact
™
test drum vial label
and the control value insert indicate the acceptable
ranges for Low Level and High Level control solu-
tions. The result displayed on your meter should fall
within the acceptable range for the level of solution
you used.
If the result is within the acceptable range:
• You can feel confident that your test drum and meter
are working properly.
53
Control Testing
If the result is not within the acceptable range:
• Check the expiration date of the test drum and control
solutions. If the test drum or control solutions are
expired, dispose of them promptly.
• Make sure that the 90-day use-by period for the test
drum has not been exceeded.
• Make sure that you used the correct control solution.
• Make sure that the optical system is clean.
• Make sure that you performed the test within the
correct temperature range +50° F to +104° F
(+10° C to +40° C).
• Make sure that the cap was placed tightly on the test
drum vial and control solutions. The test drums and
control solutions can be damaged when they are not
capped and stored properly.
• Make sure that you followed the testing steps exactly.
• Repeat the control test with a new test strip.
If the result is still not within the acceptable range, your
blood sugar results may not be reliable. Contact Accu-
Chek Customer Care
SM
for help at 1-800-858-8072,
24 hours a day.
ACCU-CHEK Compact
™
Control Range
for Lot
0000000(50)
Control Target Range
Hi
Lo
(example)