EasyManuals Logo
Home>Agilent Technologies>Multimeter>34401A

Agilent Technologies 34401A User Manual

Agilent Technologies 34401A
242 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #122 background imageLoading...
Page #122 background image
CONFigure
Example
The following program segment shows how to use the READ? command
with CONFigure to make an externally-triggered measurement.
The program configures the multimeter for dc voltage measurements.
CONFigure does not place the multimeter in the “wait-for-trigger” state.
The READ? command places the multimeter in thewait-for-trigger”
state, takes a reading when the Ext Trig terminal is pulsed, and sends
the reading to the output buffer.
CONF:VOLT:DC 10, 0.003
TRIG:SOUR EXT
READ?
bus enter statement
CONFigure
Example
The following program segment is similar to the program above but it
uses INITiate to place the multimeter in thewait-for-trigger” state.
The INITiate command places the multimeter in thewait-for-trigger”
state, takes a reading when the Ext Trig terminal is pulsed, and sends
the reading to the multimeter’s internal memory. The FETCh? command
transfers the reading from internal memory to the output buffer.
CONF:VOLT:DC 10, 0.003
TRIG:SOUR EXT
INIT
FETC?
bus enter statement
Storing readings in memory using the INITiate command is faster
than sending readings to the output buffer using the READ? command.
The multimeter can store up to 512 readings in internal memory. If you
configure the multimeter to take more than 512 readings (using the
sample count and trigger count), and then send INITiate, a memory
error is generated.
After you execute an INITiate command, no further commands are
accepted until the measurement sequence is completed. However, if you
select TRIGger:SOURce BUS, the multimeter will accept the *TRG
command (bus trigger) or an
IEEE-488 Group Execute Trigger message.
Chapter 4 Remote Interface Reference
Simplified Programming Overview
116

Table of Contents

Other manuals for Agilent Technologies 34401A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34401A and is the answer not in the manual?

Agilent Technologies 34401A Specifications

General IconGeneral
BrandAgilent Technologies
Model34401A
CategoryMultimeter
LanguageEnglish

Related product manuals