EasyManuals Logo
Home>Agilent Technologies>Multimeter>34401A

Agilent Technologies 34401A User Manual

Agilent Technologies 34401A
242 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #224 background imageLoading...
Page #224 background image
AC Characteristics
AC Characteristics
Accuracy Specifications ± ( % of reading + % of range ) [ 1 ]
Function Range [ 3 ] Frequency
24 Hour [ 2 ]
23°C ± 1°C
1 Year
23°C ± 5°C
Temperature
CoefficientC
0°C – 18°C
28°C – 55°C
True RMS
AC Voltage
[ 4 ]
100.0000 mV
1.000000 V
to
750.000 V
3 Hz 5 Hz
5 Hz 10 Hz
10 Hz 20 kHz
20 kHz 50 kHz
50 kHz 100 kHz
100 kHz 300 kHz
[6]
3 Hz 5 Hz
5 Hz 10 Hz
10 Hz 20 kHz
20 kHz 50 kHz
50 kHz 100 kHz
[5]
100 kHz 300 kHz [6]
1.00 + 0.03
0.35 + 0.03
0.04 + 0.03
0.10 + 0.05
0.55 + 0.08
4.00 + 0.50
1.00 + 0.02
0.35 + 0.02
0.04 + 0.02
0.10 + 0.04
0.55 + 0.08
4.00 + 0.50
1.00 + 0.04
0.35 + 0.04
0.05 + 0.04
0.11 + 0.05
0.60 + 0.08
4.00 + 0.50
1.00 + 0.03
0.35 + 0.03
0.05 + 0.03
0.11 + 0.05
0.60 + 0.08
4.00 + 0.50
1.00 + 0.04
0.35 + 0.04
0.06 + 0.04
0.12 + 0.05
0.60 + 0.08
4.00 + 0.50
1.00 + 0.03
0.35 + 0.03
0.06 + 0.03
0.12 + 0.05
0.60 + 0.08
4.00 + 0.50
0.100 + 0.004
0.035 + 0.004
0.005 + 0.004
0.011 + 0.005
0.060 + 0.008
0.20 + 0.02
0.100 + 0.003
0.035 + 0.003
0.005 + 0.003
0.011 + 0.005
0.060 + 0.008
0.20 + 0.02
True RMS
AC Current
[ 4 ]
1.000000 A
3.00000 A
3 Hz 5 Hz
5 Hz 10 Hz
10 Hz 5 kHz
3 Hz 5 Hz
5 Hz 10 Hz
10 Hz 5 kHz
1.00 + 0.04
0.30 + 0.04
0.10 + 0.04
1.10 + 0.06
0.35 + 0.06
0.15 + 0.06
1.00 + 0.04
0.30 + 0.04
0.10 + 0.04
1.10 + 0.06
0.35 + 0.06
0.15 + 0.06
1.00 + 0.04
0.30 + 0.04
0.10 + 0.04
1.10 + 0.06
0.35 + 0.06
0.15 + 0.06
0.100 + 0.006
0.035 + 0.006
0.015 + 0.006
0.100 + 0.006
0.035 + 0.006
0.015 + 0.006
Additional Low Frequency Errors ( % of reading )
Frequency
AC Filter
Slow Medium Fast
0 0.74 ––
0 0.22 ––
0 0.06 0.73
0 0.01 0.22
0 0 0.18
0 0 0
10 Hz 20 Hz
20 Hz 40 Hz
40 Hz 100 Hz
100 Hz 200 Hz
200 Hz 1 kHz
> 1 kHz
Sinewave Transfer Accuracy ( typical )
Frequency
10 Hz 50 kHz
50 kHz 300 kHz
Error ( % of range )
0.002%
0.005%
Conditions:
Sinewave input.
Within 10 minutes and ± 0.5°C.
Within ±10% of initial voltage and ±1% of initial frequency.
Following a 2-hour warm-up.
Fixed range between 10% and 100% of full scale ( and <120 V ).
Using 6
1
2
digit resolution.
Measurements are made using accepted metrology practices.
90 Day
23°C
± 5°C
Additional Crest Factor Errors ( non-sinewave ) [ 7 ]
Crest Factor
1 2
2 3
3 4
4 5
Error ( % of reading )
0.05%
0.15%
0.30%
0.40%
Chapter 8 Specifications
218

Table of Contents

Other manuals for Agilent Technologies 34401A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34401A and is the answer not in the manual?

Agilent Technologies 34401A Specifications

General IconGeneral
BrandAgilent Technologies
Model34401A
CategoryMultimeter
LanguageEnglish

Related product manuals