Appendix B 233
Information for Replacing 4278A with 4288A
Example of Replacing Major Functions (comparison of programs)
B. Information for Replacing
4278A with 4288A
Tolerance mode
Table B-19 compares the programs used by the 4278A and 4288A to set the limit ranges.
These ranges are the Cp limit range as shown in the following figure and the D limit range
from 0 to 0.1 in the tolerance mode.
Table B-19 Comparison of programs used to set limit ranges in tolerance mode
4278A 4288A
For the % tolerance mode:
10 ASSIGN @Agt4278a TO 717
20 !
30 OUTPUT @Agt4278a;"MPAR1"
40 OUTPUT @Agt4278a;"LPAR1"
50 !
60 OUTPUT @Agt4278a;"CLIM"
70 OUTPUT @Agt4278a;"LMOD1"
80 OUTPUT @Agt4278a;"NOM=1.0E-9"
90 OUTPUT @Agt4278a;"BIN1=-1,1"
100 !
110 OUTPUT @Agt4278a;"BIN2=-2,2"
120 !
130 OUTPUT @Agt4278a;"BIN3=-3,3"
140 !
150 !
160 !
170 !
180 OUTPUT @Agt4278a;"SLIM=0,0.1"
190 !
200 !
210 END
For the % tolerance mode:
10 ASSIGN @Agt4288a TO 717
20 !
30 OUTPUT @Agt4288a;":CALC1:FORM CP"
40 OUTPUT @Agt4288a;":CALC2:FORM D"
50 !
60 !
70 OUTPUT @Agt4288a;":CALC:COMP:MODE PCNT"
80 OUTPUT @Agt4288a;":CALC:COMP:PRIM:NOM 1.0E-9"
90 OUTPUT @Agt4288a;":CALC:COMP:PRIM:BIN1 -1,1"
100 OUTPUT @Agt4288a;":CALC:COMP:PRIM:BIN1:STAT ON"
110 OUTPUT @Agt4288a;":CALC:COMP:PRIM:BIN2 -2,2"
120 OUTPUT @Agt4288a;":CALC:COMP:PRIM:BIN2:STAT ON"
130 OUTPUT @Agt4288a;":CALC:COMP:PRIM:BIN3 -3,3"
140 OUTPUT @Agt4288a;":CALC:COMP:PRIM:BIN3:STAT ON"
150 FOR I=4 TO 9
160 OUTPUT @Agt4288a;":CALC:COMP:PRIM:BIN"&VAL$ (I)&":STAT OFF"
170 NEXT I
180 OUTPUT @Agt4288a;":CALC:COMP:SEC 0,0.1"
190 OUTPUT @Agt4288a;":CALC:COMP:SEC:STAT ON"
200 !
210 END
For the absolute tolerance mode:
10 ASSIGN @Agt4278a TO 717
20 !
30 OUTPUT @Agt4278a;"MPAR1"
40 OUTPUT @Agt4278a;"LPAR1"
50 !
60 OUTPUT @Agt4278a;"CLIM"
70 OUTPUT @Agt4278a;"LMOD2"
80 OUTPUT @Agt4278a;"NOM=1.0E-9"
90 OUTPUT @Agt4278a;"BIN1=-1.0E-11,1.0E-11"
100 !
110 OUTPUT @Agt4278a;"BIN2=-2.0E-11,2.0E-11"
120 !
130 OUTPUT @Agt4278a;"BIN3=-3.0E-11,3.0E-11"
140 !
150 !
160 !
170 !
180 OUTPUT @Agt4278a;"SLIM=0,0.1"
190 !
200 !
210 END
For the absolute tolerance mode:
10 ASSIGN @Agt4288a TO 717
20 !
30 OUTPUT @Agt4288a;":CALC1:FORM CP"
40 OUTPUT @Agt4288a;":CALC2:FORM D"
50 !
60 !
70 OUTPUT @Agt4288a;":CALC:COMP:MODE DEV"
80 OUTPUT @Agt4288a;":CALC:COMP:PRIM:NOM 1.0E-9"
90 OUTPUT @Agt4288a;":CALC:COMP:PRIM:BIN1 -1.0E-11,1.0E-11"
100 OUTPUT @Agt4288a;":CALC:COMP:PRIM:BIN1:STAT ON"
110 OUTPUT @Agt4288a;":CALC:COMP:PRIM:BIN2 -2.0E-11,2.0E-11"
120 OUTPUT @Agt4288a;":CALC:COMP:PRIM:BIN2:STAT ON"
130 OUTPUT @Agt4288a;":CALC:COMP:PRIM:BIN3 -3.0E-11,3.0E-11"
140 OUTPUT @Agt4288a;":CALC:COMP:PRIM:BIN3:STAT ON"
150 FOR I=4 TO 9
160 OUTPUT @Agt4288a;":CALC:COMP:PRIM:BIN"&VAL$ (I)&":STAT OFF"
170 NEXT I
180 OUTPUT @Agt4288a;":CALC:COMP:SEC 0,0.1"
190 OUTPUT @Agt4288a;":CALC:COMP:SEC:STAT ON"
200 !
210 END

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4288A and is the answer not in the manual?

Agilent Technologies 4288A Specifications

General IconGeneral
BrandAgilent Technologies
Model4288A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals